Trace Chlorine dioxide CCS241

This product is no longer available.

Please check new generation in the section predecessor / successor or contact your local Endress+Hauser entity.

Funkcje i specyfikacja

Powrót

Details

  • Zasada pomiaru

    Chlorine dioxide

  • Aplikacja

    Drinking water, water, process

  • Charakterystyka

    Amperometric measurement of dissolved chlordioxide

  • Zakres pomiarowy

    0,01 - 5ppm Chlordioxide

  • Zasada działania

    - closed (membrane covered) measuring cell.
    - Reduction of chlordioxide (ClO2) to chlorid at the cathode with appr. 120mV.

  • Konstrukcja

    Closed amperometric 2-electrode measuring cell with PTFE membrane.

  • Materiał

    Sensor shaft : PVC
    Membrane : PTFE
    Membrane cap : PBT (GF30); PVDF

  • Wymiar

    Diameter : appr. 25mm
    Length : 160mm

  • Temperatura procesu

    2°C ... 45°C

  • Ciśnienie procesu

    0bar - 1bar

  • czujnik temperatury

    With NTC-temperature sensor

  • Podłączenie

    Fixed cable

Dokumentacja / Instrukcje obsługi / Oprogramowanie

Akcesoria / Części zamienne

Nr
Order code
Specification
  • 1
  • 2
  • A
    • Order code 50005181

      Junction box VBC

      Przy przedłużaniu przewodów CMK. Stopień ochrony: IP65

    • Order code 50085333

      Kabel pomiarowy CYK 71

      Do elektrod pH/Redox oraz dwuelektrodowych cel konduktometrycznych. Minimalna długość 5 m

    • Order code 71257946

      Fotometr Macherey-Nagel PF3 (chlor)

      Fotometr do pomiarów wolnego chloru oraz chloru całkowitego w zakresie 0.05 - 6 mg/l. Aplikacje: czysta woda oraz ścieki. Zawiera walizkę, kuwety, fotometr oraz reagenty zgodnie z metodą DPD (ISO 7393).

    • Order code COY8-

      Żel "0"(COY8) do kalibracji/weryfikacji

      Żel, do sprawdzania punktu zerowego: Weryfikacja, kalibracja, dostrojenie dla sond tlenowych i czujników do dezynfekcji ::Amperometrycznych i optycznych ::o średnicy 12, 25 i 40 mm ::aplikacja w laboratorium i w terenie

  • Następca